Specifications |
Temperature Range
|
-30 °C to 650 °C (-22 °F to 1202 °F) |
Accuracy |
±1.0 °C or ±1.0% of reading, whichever is greater
-10 °C to 0 °C: ±2.0
-30 °C to -10 °C: ±3.0 |
Response Time (95%) |
Fluke 62 MAX+ |
<300 ms (95% of reading) |
Spectral Response |
8 to 14 microns |
Emissivity |
0.10 to 1.00 |
Optical Resolution
|
12:1 (calculated at 90% energy) |
Display Resolution |
0.1 °C (0.2 °F) |
Repeatability of Readings |
±0.5% of reading or <±0.5 °C (1 °F), whichever is greater |
Power |
AA battery |
Battery Life |
8 hours with laser and backlight on |
Physical Specifications |
Weight |
255 g (8.99 oz) |
|
Size |
(175 x 85 x 75) mm (6.88 x 3.34 x 2.95) inches |
|
Operating Temperature |
0 °C to 50 °C (32 °F to 122 °F) |
|
Storage Temperature |
-20 °C to 60 °C (-4 °F to 140 °F), (without battery) |
|
Operating Humidity |
10 % to 90 % RH non-condensing at 30 °C (86 °F) |
|
Operating Altitude |
2000 meters above mean sea level |
|
Storage Altitude |
12,000 meters above mean sea level |
|
IP Rating |
IP 54 per IEC 60529 |
|
Drop Test |
3 meters |
|
Vibration and Shock |
IEC 68-2-6 2.5 g, 10 to 200 Hz, IEC 68-2-27, 50 g, 11 ms |
|
EMC |
EN 61326-1:2006 EN 61326-2:2006 |
Standards and Agency Approval
|
Compliance |
EN/IEC 61010-1: 2001 |
|
Laser Safety |
FDA and EN 60825-1 Class II |