Fluke 62 MAX+ Infrared Thermometer
Specifications |
Temperature Range
|
-30 °C to 650 °C (-22 °F to 1202 °F) |
Accuracy |
±1.0 °C or ±1.0% of reading, whichever is greater
-10 °C to 0 °C: ±2.0
-30 °C to -10 °C: ±3.0 |
Response Time (95%) |
Fluke 62 MAX+ |
<300 ms (95% of reading) |
Spectral Response |
8 to 14 microns |
Emissivity |
0.10 to 1.00 |
Optical Resolution
|
12:1 (calculated at 90% energy) |
Display Resolution |
0.1 °C (0.2 °F) |
Repeatability of Readings |
±0.5% of reading or <±0.5 °C (1 °F), whichever is greater |
Power |
AA battery |
Battery Life |
8 hours with laser and backlight on |
Physical Specifications |
Weight |
255 g (8.99 oz) |
Size |
(175 x 85 x 75) mm (6.88 x 3.34 x 2.95) inches |
Operating Temperature |
0 °C to 50 °C (32 °F to 122 °F) |
Storage Temperature |
-20 °C to 60 °C (-4 °F to 140 °F), (without battery) |
Operating Humidity |
10 % to 90 % RH non-condensing at 30 °C (86 °F) |
Operating Altitude |
2000 meters above mean sea level |
Storage Altitude |
12,000 meters above mean sea level |
IP Rating |
IP 54 per IEC 60529 |
Drop Test |
3 meters |
Vibration and Shock |
IEC 68-2-6 2.5 g, 10 to 200 Hz, IEC 68-2-27, 50 g, 11 ms |
EMC |
EN 61326-1:2006 EN 61326-2:2006 |
Standards and Agency Approval
|
Compliance |
EN/IEC 61010-1: 2001 |
Laser Safety |
FDA and EN 60825-1 Class II |
Fluke T+PRO Electrical Testers
Specifications |
Temperature |
Operating |
-10 °C to 55 °C (14 °F to 131 °F) |
Storage |
-30 °C to 60 °C (-22 °F to 140 °F) |
Relative humidity |
0 °C to 30 °C (32 °F to 86 °F): 90 % |
30 °C to 40 °C (86 °F to 104 °F): 75 % |
40 °C to 50 °C (104 °F to 122 °F): 45 % |
Altitude |
Operating |
2,000 m (approx. 6,561.6 ft) |
Storage |
10,000 m (approx. 32,808.4 ft) |
Battery type/life |
AAA (2); 40 hours |
Shock, vibration |
Sinusoidal vibration per MIL-PRF-28800F for a Class 2 instrument |
Safety |
ISA-82.02.01 (IEC 61010-1 Mod) Second Edition, CAN/CSA-C22.2 No. 61010-1 Second Edition, IEC 61010 - 1 to 1000 V CAT III/600 V CAT IV |
Compliance to NFPA 70E* |
AC Bandwidth |
45 Hz to 66 Hz |
Maximum working voltage |
1000 V ac/dc |
Maximum measurable voltage |
600 V ac/dc |
Duty cycle |
Indefinitely for voltages up to 240 V |
For voltages between 240 V and 600 V the duty cycle is 30 s on / 300 s off |
For voltages above 240 V the tester must connect to a voltage source only for a MAXIMUM of 30 s and then disconnect for a MINIMUM of 300 s |
Voltage hazard LED |
LEDs turn on at voltages > 30 V ac/dc ± 35 % |
LED voltage indication levels |
12 V, 24 V, 48 V, 120 V, 208 V, 240 V, 277 V, 480 V, 600 V |
LED voltage indication levels for Canadian version |
12 V, 24 V, 48 V, 120 V, 208 V, 240 V, 347 V, 480 V, 600 V |
LEDs |
LEDs turn on between 90 % and 100 % of the indicated voltage on the LED, except for the 12 V LED which turns on between 50 % and 100 % |
LCD Accuracy on T+PRO only |
AC voltage |
± (3 % rdg + 2 digits) |
DC voltage |
± (2 % rdg + 2 digits) |
Resistance |
± (5 % rdg + 3 digits) |
LCD Resolution |
0.1 V for voltages < 50 V, 1 V for voltages ≥ 50 V
0.01 kΩ for resistance measurement (For T+PRO Only) |
GFCI Test current |
100 V - 150 V @ 6 mA - 9 mA ac, 150 V - 240 V < 12 mA |
Standard input test current |
< 5 mA |
Size (HxWxL) |
1.3 in x 2.15 in x 7.6 in (3.3 cm x 5.46 cm x 19.3 cm ) |
Weight |
9.9 oz (0.28 kg) |
*When used properly per Article 110.9 Use of Equipment (A) Test Instruments and Equipment.
Fluke 1AC-II VoltAlert
Specifications |
Operating principle |
Senses the steady state electrostatic field produced by ac voltage through insulation without requiring contact to the bare conductor. A red glow at the tip and a beeping noise (if not switched OFF) indicates the presence of voltage. |
Voltage sensing ranges |
Nominally, 90 V AC to 1000 V AC or 200 V AC to 1000 V AC depending on model, 45 Hz to 405 Hz; also a 20 V to 90 V control circuit model |
Detector tip style |
Blade or round, depending on model |
Light source |
One high intensity red LED |
Agency Approvals |
c CSA us, CE, C-Tick |
Safety Rating |
1000 V, CAT IV |
IP Rating |
IP 40 |
Operating Temperature |
-10 °C to 50 °C |
Operating humidity ranges |
0 % to 95 % |
(0 °C to 30 °C) |
0 % to 75 % |
(30 °C to 40 °C) |
0 % to 45 % |
(40 °C to 55 °C) |
|
Operating Altitude |
3000 meters |
Batteries |
2 AAA alkaline batteries included |
Note: Product does not respond to random electrostatic discharge. |